This feature improves control over Quality Control (QC) values coming from devices, by introducing a review workflow before data is plotted on LJ charts.
It helps lab teams ensure that only verified QC data is visualized and used for monitoring instrument performance.
Feature Overview
What Is It?
QC Result-wise Validation allows labs to:
Review each QC value before it reflects on the LJ chart
See a summary of QC results (Validated vs Pending) per device
Receive alerts when results are pending review
Set rules lab-wide or per-device for how QC data should flow
How It Works – End-to-End Flow
Step | Action |
---|---|
1 | Device sends a QC result |
2 | If review is enabled, result lands in the QC Tab (status: Pending Review ) |
3 | Reviewer can Release (approve) or Cancel (reject) |
4 | Once Released, the result appears in the LJ Chart |
5 | Until reviewed, the chart shows a warning banner: Some QC values are pending review. Please review them to reflect in the LJ chart. |
Where Does It Show Up?
Waiting list->Device Grid View:
Shows summary counts for Validated and Pending QC results for each device.Click on the device
“QC” Tab (next to "Next Results"):
Lists incoming QC values
Allows action per result (Release / Cancel)
Video : https://drive.google.com/file/d/1y-b9ld3pnf_nFm-G1CZ7cV6UkQ7Q8Urf/view?usp=sharing
Settings Configuration
This feature requires settings at two levels: lab-level and device-level.
1. Lab-Level Setting: Enable QC Result Validation
Location:Operations Dashboard > Waiting List Settings > Lab Workflow Settings
Setting Name:
Device Result Validation
Default: OFF
Visible To: Admins and Support Dashboard Users only
2. Device-Level Setting: Manual Review Toggle
Location:Device Settings > Device Validation Settings
Toggle:
Manually review QC values before adding to LJ
When enabled:
QC results go to QC Tab for review
Are excluded from LJ chart until released